Digital Photo Colorimeter

Digital Photo Colorimeter

Photo Colorimeter

This is ready to use benchtop instrument especially designed to greater degree of efficient performance of photo colorimeter analysis of a concentration in laboratory. This photo colorimeter is a compact instrument, occupies low bench space and feature many user-friendly controls enabling users to operate the machine with simplicity.

Analysis output can be seen on LED display in terms of %Transmission (%T), Absorbance (Abs.), Concentration (Conc.), and K Factor. Imported glass filters are used that cover wavelength from 400 nm to 700 nm and provide superior quality results. Filter settings are adjusted by rotating disc knob having 8 filters. Automatic shutter keeps the photocell safe from light when not in use, this increases service life of photocell. This digital photo colorimeter is equipped with printer facility and excellent memory storage up to 100 samples.

This unit has in-built function of saving the last K factor and the standard concentration value. 10 mm path length matched test tubes are used for samples. User can also enter any known 'Factor' for a test.

Features
400nm to 700nm wavelength Compact design low bench space
Excellent memory up to 100 samples Microprocessor circuit for precise measurements
Long service life photocell Automatic shutter keeps the photocell light safe
Printer facility Supplied as compete test kit

Specifications: PX/PCM/1312
Model No. PX/PCM/1312
Wave Length Range 400 to 700nm
Resolution 1% T, 0.01 Abs., 1 Conc., 1 K Factor
% Transmission 0 - 100% T
Absorbance 0 - 1.99 A
Concentration 0- 19999
K Factor 1 - 19999
Photo Detector Silicon Photodiode/Photo Cell
Display 16x2 line alphanumeric LCD display with backlit for %T, Abs., Conc., K Factor
Keyboard 8 Keys, soft touch membrane type
Data Storage Upto 100 samples
Light Source 6.8V, 0.3Amp. Tungsten Lamp
Sample System 10 mm path length matched glass test tubes
Filters 420, 440, 490, 520, 540, 570, 600, 700nm
Printer Interface Centronics Parallel Printer interface for any dot matrix printer
RS232C Interface Available (on request)
Dimensions 240 x 254 x 90 mm (L x B x H)
Weight 2.5 Kg. (Approx.)
Power Supply 230V AC+10% 50Hz. with in-built stabilizer
Accessories Matched Test Tube Set of 5, Operation Manual, Dust Cover, Spare Lamp

Specifications: PX/PCM/312, PX/PCM/313
Model No. PX/PCM/312 PX/PCM/313
High Std. Glass Filters 8 Filters 5 Filters
Minimum Volume 1 ml
Display 2 ½ digit LED
Range 400 to 700 nm
Output OD : 0 to 1.99
Resolution OD : 0.01
Accuracy 0.5% FS ±1 Digit
Dectector Photo Cell/Photo Diode
light source (tungston lamp) 6.8 V, 300 mA
Power 230 V ± 10% AC, 50Hz
Dimensions 195 X 235 X 105 mm
Weight 2 Kgs. (Approx.)
Filters 8 filters - 400 450 490 520 540 570 620 680 nm
5 filters - 450 490 540 620 680 nm

Specifications: PX/PCM/1311, PX/PCM/1313
Model No. PX/PCM/1311 PX/PCM/1313
Wave Length Range 400 to 700nm
Display 3 Digit 7 Segment Red LED Display
Resolution 1% T, 0.01 Abs., 1% T, 0.01 Abs., Conc. 1,
K Factor 1.
Transmittance (%T) 0-100% 0-100%
Absorbance 0-1.99A 0-1.99A.
Concentration - 1.0-999
K Factor - 1.0-999
Key Board(Soft Touch Member Type) 2 Keys, 3 Keys,
Autozeroing Available Available
Light Source 6.3 V 0.3 amp Tungsten Lamp
solution volume 1ml (Minimum)
Sample System 10mm path length matched glass test tubes
Optical Filters 8 filter mounted on the turret
400, 450, 490, 520, 540, 570, 620, 680 nm
Weight 2½ Kg. Approx.
Power Supply 230 VAC to + 10% 50Hz. With in - built stabilizer
Accessories Spare Lamp, Test Tubes set (Matched),
Dust Cover, Instruction Manual